2018
DOI: 10.1186/s13014-018-1106-y
|View full text |Cite
|
Sign up to set email alerts
|

Validation of new transmission detector transmission factors for online dosimetry: an experimental study

Abstract: BackgroundThe demand for dose verification during treatment has risen with the increasing use of intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) in modern radiation therapy. This study aims to validate the transmission factors of a new transmission detector, the Dolphin online monitoring system (IBA Dosimetry, Schwarzenbruck, Germany), for clinical use.MethodsThe transmission factors of the Dolphin detector were evaluated using 6 MV, 6 flattening filter free (FFF), 10 M… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
8
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 17 publications
(8 citation statements)
references
References 30 publications
0
8
0
Order By: Relevance
“…Attenuation of 6 MV beams along the central axis for the various detectors they compared ranged from 1.0% to 7.2%, with ion-chamberbased devices typically having higher attenuation (ranging from 3.3% to 7.2%) and diode transmission devices typically having lower attenuation (~1%). Another available transmission detector not included in the comparison by Li et al [13] is the Dolphin detector (IBA Dosimetry, Schwarzenbuck, Germany) [18,20,21]. The Dolphin is a 2D array of 1513 plane-parallel ion chambers.…”
Section: Discussionmentioning
confidence: 99%
“…Attenuation of 6 MV beams along the central axis for the various detectors they compared ranged from 1.0% to 7.2%, with ion-chamberbased devices typically having higher attenuation (ranging from 3.3% to 7.2%) and diode transmission devices typically having lower attenuation (~1%). Another available transmission detector not included in the comparison by Li et al [13] is the Dolphin detector (IBA Dosimetry, Schwarzenbuck, Germany) [18,20,21]. The Dolphin is a 2D array of 1513 plane-parallel ion chambers.…”
Section: Discussionmentioning
confidence: 99%
“…This small variation could be the result of the simple homogeneous design of the IQM detector. For the dolphin transmission detector, the ratio of the TF of the 5 cm × 5 cm and the 20 cm × 20 cm field was reported to be 7.6% …”
Section: Discussionmentioning
confidence: 99%
“…A very similar approach has been used by Park et al for the dolphin transmission detector. Furthermore, the obtained dTF was compared to the TF of regular open fields.…”
Section: Methodsmentioning
confidence: 99%
“…[10][11][12] PD has been established utilizing EPID for pretreatment verification. 13,14) 138 www.ksmp.or.kr Modern radiotherapy has increased the need for online dose verification owing to the complexity and accuracy of the new delivery techniques. (e.g., VMAT or Stereotactic ablative radiotherapy (SABR)).…”
Section: ) Portal Dosimetry (Pd)mentioning
confidence: 99%