Formulas to Explain Popular Oscillometric Blood Pressure Estimation Algorithms

Abstract: Oscillometry is the blood pressure (BP) measurement principle of most automatic cuff devices. The oscillogram (which is approximately the blood volume oscillation amplitude-external pressure function) is measured, and BP is then estimated via an empirical algorithm. The objective was to establish formulas to explain three popular empirical algorithms in the literature—the maximum amplitude, derivative, and fixed ratio algorithms. A mathematical model of the oscillogram was developed and analyzed to derive para… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
32
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
6
3

Relationship

0
9

Authors

Journals

citations
Cited by 50 publications
(32 citation statements)
references
References 26 publications
(60 reference statements)
0
32
0
Order By: Relevance
“…sBP init and dBP init retrieved from finger oscillometry will need a heart level correction sensor as well as an adapted transfer function known from literature 29,37 . The same is true for finger oscillometric algorithms including its parametrization 38 , although a detailed development work including validation will be needed.…”
Section: Resultsmentioning
confidence: 99%
“…sBP init and dBP init retrieved from finger oscillometry will need a heart level correction sensor as well as an adapted transfer function known from literature 29,37 . The same is true for finger oscillometric algorithms including its parametrization 38 , although a detailed development work including validation will be needed.…”
Section: Resultsmentioning
confidence: 99%
“…Proprietary algorithms are then applied to detect mean arterial pressure, systolic BP and diastolic BP. Mean arterial pressure is conventionally assumed to be the peak amplitude of the oscillometric envelope, but because algorithms are proprietary, the precise method of estimation of systolic and diastolic BP is usually undisclosed [12,13].…”
Section: Cuff Blood Pressure Measurement Methodsmentioning
confidence: 99%
“…The upper and lower envelopes of the oscillation are traced as cuff pressure varies from above systolic pressure (SP) to below diastolic pressure (DP). Algorithms then estimate the mean, systolic, and diastolic pressures from the oscillogram (plot of the oscillation amplitude vs. cuff pressure) ( 30 , 31 ).…”
Section: Non-invasive Smartphone Contact Measurements Of Blood Pressurementioning
confidence: 99%