2010
DOI: 10.1016/j.jpba.2010.06.022
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of the Architect® tacrolimus assay in kidney, liver, and heart transplant recipients

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

3
22
0

Year Published

2011
2011
2023
2023

Publication Types

Select...
7
1
1

Relationship

0
9

Authors

Journals

citations
Cited by 46 publications
(29 citation statements)
references
References 21 publications
3
22
0
Order By: Relevance
“…Immunoassay results were overestimated compared to LC-MS/MS by a mean of 18.1% for tacrolimus, 41.4% for sirolimus, and 15.6% for cyclosporine ( Fig. 3D-F), which is comparable to previously published biases for the Abbott Architect chemiluminescent immunoassays [22][23][24][25]. …”
Section: Correlation With Immunoassayssupporting
confidence: 85%
“…Immunoassay results were overestimated compared to LC-MS/MS by a mean of 18.1% for tacrolimus, 41.4% for sirolimus, and 15.6% for cyclosporine ( Fig. 3D-F), which is comparable to previously published biases for the Abbott Architect chemiluminescent immunoassays [22][23][24][25]. …”
Section: Correlation With Immunoassayssupporting
confidence: 85%
“…After an overnight fast, patients' morning tacrolimus dose was administered in capsules, and serial blood samples were drawn over the course of the dosing interval at times 0 (predose), and 0.5, 1, 2, 4, 6, 8, and 12 h postdose. EDTA whole-blood samples were analyzed either fresh at room temperature or after being frozen at Ϫ80°C via whole-blood immunoassay [Architect tacrolimus in whole blood (Bazin et al, 2010), Mayo Clinic Arizona, Phoenix, AZ].…”
Section: Methodsmentioning
confidence: 99%
“…6,9,11) The 20% CV values for ACMIA on 19 different Dimension ® analyzers in the present study were slightly lower than the reported LOQ values. 6,9,11) Although the linearity for ACMIA is reported to be up to at least 31.4 ng/ mL, 11) ACMIA tended to have a negative bias at higher tacrolimus concentrations. The present results completely agree with the previous report using spiked samples by Levine et al 12) In addition, Hashi et al 13) also reported that the ME for ACMIA at tacrolimus concentrations greater than 10 ng/mL has a more negative bias.…”
Section: Discussionmentioning
confidence: 96%
“…LOQ values for CLIA and ECLIA are reported to be 0.5-0.8 [6][7][8][9] and 0.2-0.3 ng/mL, 10) respectively. These LOQ values were calculated by Architect ® analyzer or cobas ® analyzer On the other hand, the LOQ value for ACMIA is reported to be 2.5-3.12 ng/mL.…”
Section: Discussionmentioning
confidence: 99%