Different configurations of photoacoustic (PA) setups for the online-measurement of gaseous NO, employing semiconductor lasers at 2.9 and 4.5 μm, were developed and tested. Their performance was assessed with respect to the analysis of NO emissions from wastewater treatment plants. For this purpose, the local NO emissions of a wastewater treatment bioreactor was sampled by a dedicated mobile sampling device, and the total NO emissions were analyzed in the gastight headspace of the bioreactor. We found that the use of a quantum-cascade laser emitting at about 4.53 μm, operated in a wavelength modulation mode, in combination with a conventional longitudinal PA cell yielded the highest sensitivity (<100 ppbv). However, we also observed a strong cross-sensitivity to humidity, which can be explained by increased V-T relaxation. This observation in combination with the limited dynamic range (max conc. ∼ 3000 ppmv) led us to the use of the less-sensitive but spectroscopically more robust 2.9 μm laser. A detection limit below 1 ppmv, a dynamic range of more than 4 orders of magnitude, no influence of humidity or any other substance relevant to the off-gas analysis, as well as a comparable low price of the laser source made it the ideal tool for NO analyses of the off-gas of a wastewater treatment plant. Such a system was implemented successfully in a full-scale wastewater treatment plant. The results regarding the comparison of different PA setups can be transferred to other systems, and the optimum performance can be selected according to the specific demands.