Evaluation of IVAC® Variable Pressure Volumetric Pump Model 560 for the Delivery of Red Blood Cells, Adenine–Saline Added

Abstract: Mechanical pump systems for the delivery of intravenous solutions have been used for the transfusion of red blood cells. In this study, the IVAC Variable Pressure Volumetric Pump Model 560 was evaluated for that purpose using flow rates of 70 mL/hour and 999 mL/hour through 16-, 19-, and 23-gauge needles and an 18-gauge angiocatheter. The largest degree of hemolysis induced by the delivery system resulted in a mean increase of plasma hemoglobin of 150 mg/L (15 mg/dL); this is equivalent to the loss of 0.03% of… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
11
0

Year Published

1993
1993
2017
2017

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 12 publications
(11 citation statements)
references
References 0 publications
0
11
0
Order By: Relevance
“…Four publications (23.5%) noted significant statistical alterations in the outcome variables. However they opted on determining the presence of hemolysis being clinically significant for values where the level of hemolysis was above 0.8% 18 , 20 , 25 , 28 .…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…Four publications (23.5%) noted significant statistical alterations in the outcome variables. However they opted on determining the presence of hemolysis being clinically significant for values where the level of hemolysis was above 0.8% 18 , 20 , 25 , 28 .…”
Section: Resultsmentioning
confidence: 99%
“…No publication stated the value of the infusion pressure as a possible factor for cellular damage. They only stated the pressor variation in the equipment in high infusion fluxes 20 . Aspects related to accessories to infusion pumps such as catheters, were described and evaluated only in conjunction with the infusion system.…”
Section: Discussionmentioning
confidence: 99%
“…Most of the infusion pumps previously tested with packed RBCs seemed to induce relatively low levels of haemolysis . However, there were some notable exceptions, including the pump that was previously used at our institution .…”
Section: Discussionmentioning
confidence: 98%
“…Most of the infusion pumps previously tested with packed RBCs seemed to induce relatively low levels of haemolysis [12][13][14][15][16][17][18][19][20][21][22]. However, there were some notable exceptions, Change in difference between sample before and after pump utilization without the antisiphon valve: S6-(S1 + S2) /2.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation