Evaluation of Four Automated Methods for Determination of Whole Blood Cyclosporine Concentrations

Abstract: Cyclosporine is a widely used and potent immunosuppressant drug with a narrow therapeutic index. Therefore, cyclosporine concentrations should be monitored closely. Various automated immunologic methods for cyclosporine whole blood determinations are available. Two new methods, fluorescence polarization immunoassay (FPIA) for the AxSYM by Abbott Laboratories, Chicago, IL, and the cloned enzyme donor immunoassay (CEDIA) by Boehringer Mannheim, Mannheim, Germany, have been introduced. In addition, Dade Behring i… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
15
0

Year Published

2005
2005
2014
2014

Publication Types

Select...
5
3
2

Relationship

0
10

Authors

Journals

citations
Cited by 36 publications
(16 citation statements)
references
References 0 publications
1
15
0
Order By: Relevance
“…Accuracy was 96% (400 mg l -1 ) and precision in the range applied (80-800 mg l -1 CsA) was described by the polynomial function: standard deviation = 1.2009 + 0.1278 3 . Compared with other immunoassays the FPIA (AxSYM) has low crossreactivity towards CsA metabolites: 5.5% (AM1), 13.7% (AM9), 2.1% (AM4n) and 2.5% (AM19) [33].…”
Section: Drug Analysismentioning
confidence: 99%
“…Accuracy was 96% (400 mg l -1 ) and precision in the range applied (80-800 mg l -1 CsA) was described by the polynomial function: standard deviation = 1.2009 + 0.1278 3 . Compared with other immunoassays the FPIA (AxSYM) has low crossreactivity towards CsA metabolites: 5.5% (AM1), 13.7% (AM9), 2.1% (AM4n) and 2.5% (AM19) [33].…”
Section: Drug Analysismentioning
confidence: 99%
“…The mean cyclosporine concentrations have [41][42][43][44][45][46] The FPIA assay for the application on the TDx analyzer (Abbott Laboratories) is no longer commercially available. More recently, Abbott Laboratories have marketed a CMIA for application on the Architect analyzer.…”
Section: Issues Of Interferences In Monitoring Of Immunosuppressantsmentioning
confidence: 99%
“…Many studies have been addressed specificity with regard to measurements of cyclosporine levels. [4][5][6][7][8] Recent studies have analyzed cyclosporine using HPLC-MS/ MS. [9][10][11][12] Also, some reports have been published on the simultaneous analysis of several immunosuppressant drugs (sirolimus, everolimus, and tacrolimus) [13][14][15][16][17] or the simultaneous analysis of cyclosporine metabolites. [18][19][20][21] These studies are time-consuming, and require many steps for sample preparation, including evaporation or reconstitution.…”
Section: Introductionmentioning
confidence: 99%