2010
DOI: 10.1016/j.diagmicrobio.2009.12.021
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of a new third-generation ARCHITECT rHTLV-I/II assay for blood screening and diagnosis

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

2
10
0
7

Year Published

2017
2017
2019
2019

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 12 publications
(19 citation statements)
references
References 43 publications
2
10
0
7
Order By: Relevance
“…These findings are consistent with previously reported results. In an earlier study of 406 precharacterized HTLV-1/2-positive samples, Abbott Architect rHTLV-I/II had a sensitivity of 100% (21, 22). Similarly, in a study of 714 confirmed HTLV-1/2-positive samples, a sensitivity of 100% was reported for Abbott Prism HTLV-I/HTLV-II (23).…”
Section: Discussionmentioning
confidence: 93%
See 1 more Smart Citation
“…These findings are consistent with previously reported results. In an earlier study of 406 precharacterized HTLV-1/2-positive samples, Abbott Architect rHTLV-I/II had a sensitivity of 100% (21, 22). Similarly, in a study of 714 confirmed HTLV-1/2-positive samples, a sensitivity of 100% was reported for Abbott Prism HTLV-I/HTLV-II (23).…”
Section: Discussionmentioning
confidence: 93%
“…The results for the comparator assays were consistent with previous data. In a study of 5,646 blood donor samples, Abbott Architect rHTLV-I/II had a specificity of 99.95%, compared with 99.86% when tested in 692 routine diagnostic samples (21, 22). Ortho Avioq HTLV-I/II had a specificity of 99.95% in a study of 11,415 blood donor samples (24), while Abbott Prism HTLV-I/HTLV-II had a specificity of 99.93% in a study of 21,943 blood donor samples (23).…”
Section: Discussionmentioning
confidence: 99%
“…From the third generation, the recombinant proteins and synthetic peptides were used to capture antibodies by double-antigen sandwich method. 11 These technological advances have demonstrated that the previous comparative assessments showed higher sensitivity and specificity with the third generation than with the first and second generation. [12][13][14] F I G U R E 1 Comparison of the two anti-HTLV assays, Elecsys HTLV-I/II and ARCHITECT rHTLV-I/II, with proven anti-HTLV-positive specimens (n = 100) F I G U R E 2 Frequency histogram for the S/CO results obtained from the potential blood donor screening (n = 500).…”
Section: Discussionmentioning
confidence: 99%
“…For the first‐ and second‐generation assays, an indirect format was applied using viral lysate as an antigen. From the third generation, the recombinant proteins and synthetic peptides were used to capture antibodies by double‐antigen sandwich method . These technological advances have demonstrated that the previous comparative assessments showed higher sensitivity and specificity with the third generation than with the first and second generation …”
Section: Discussionmentioning
confidence: 99%
“…Recently, a variety of sensitive diagnostic kits for serological assays have become available to detect both HTLV-1 and HTLV-2 infections. [9][10][11] However, because pathogenesis between HTLV-1 and HTLV-2 are largely different, discrimination is important when specimens show positivity in these tests. Western blotting is often used to confirm the infection, but indeterminate results occur frequently.…”
Section: Introductionmentioning
confidence: 99%