Green RP-HPLC method for analysis of indomethacin (IND) in bulk drug, nanoemulsion and various marketed formulations was developed and validated in present investigation.The chromatographic identification was achieved on Lichrosphere 250 X 4.0 mm RP C 8 column having a 5µm packing as a stationary phase using a green solvent ethyl acetate (100 % v/v) as a mobile phase, at a flow rate of 1.0 mL/min with UV detection at 318 nm.The proposed method was validated for linearity, selectivity, accuracy, precision, robustness, sensitivity and specificity. The utility of the method was verified by assay of IND in developed nanoemulsion, marketed tablet, capsule, suppository and injection. The proposed method was found to be selective, precise, accurate, robust, sensitive and specific. The amount of IND in developed nanoemulsion and all marketed formulations ranged from 98.34-101.40 %. The proposed method successfully resoled IND peak in the presence of its degradation products which indicated stability-indicating property of the Downloaded by [Georgetown University] at 23:38 17 September 2013 ACCEPTED MANUSCRIPT ACCEPTED MANUSCRIPT 2 proposed method. These results indicated that the proposed method can be successfully employed for routine analysis of IND in bulk drug and pharmaceutical dosage forms.