Determination of 237Np and 239Pu in urine using sector field inductively coupled plasma mass spectrometry (SF-ICP-MS)

Abstract: Measuring 237 Np and 239 Pu in urine at low levels is important for both biomonitoring and radiological emergency response. Here we report a newly developed and validated analytical method used to determine 237 Np and 239 Pu in urine by selective retention of Np and Pu from 2mL of urine directly onto TEVA® resin followed by SF-ICP-MS (coupled to a membrane desolvating introduction system) detection. The method provides solid phase extraction of Np/Pu with observed recovery ratios ranging from 89% to 113% and r… Show more