2013
DOI: 10.1063/1.4788922
|View full text |Cite
|
Sign up to set email alerts
|

Design criteria for developing low-resource magnetic bead assays using surface tension valves

Abstract: Many assays for biological sample processing and diagnostics are not suitable for use in settings that lack laboratory resources. We have recently described a simple, self-contained format based on magnetic beads for extracting infectious disease biomarkers from complex biological samples, which significantly reduces the time, expertise, and infrastructure required. This self-contained format has the potential to facilitate the application of other laboratory-based sample processing assays in lowresource setti… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
28
0

Year Published

2013
2013
2023
2023

Publication Types

Select...
9

Relationship

3
6

Authors

Journals

citations
Cited by 26 publications
(28 citation statements)
references
References 21 publications
0
28
0
Order By: Relevance
“…While this illustrates that vortexing is an excellent means of dispersing beads throughout a fluid in a microcentrifuge tube, vortexing the pre-arrayed assay tube would result in hydrodynamic instabilities that would combine the otherwise separated fluid chambers. 27 This re-iterates the importance of mixing strategies to improve the rate of target biomarker binding, and hence shorten processing times, in pre-arrayed microfluidic cassettes designed for off-theshelf use.…”
Section: Discussionmentioning
confidence: 93%
See 1 more Smart Citation
“…While this illustrates that vortexing is an excellent means of dispersing beads throughout a fluid in a microcentrifuge tube, vortexing the pre-arrayed assay tube would result in hydrodynamic instabilities that would combine the otherwise separated fluid chambers. 27 This re-iterates the importance of mixing strategies to improve the rate of target biomarker binding, and hence shorten processing times, in pre-arrayed microfluidic cassettes designed for off-theshelf use.…”
Section: Discussionmentioning
confidence: 93%
“…27 The tube format has been demonstrated as a method for tuberculosis DNA extraction and amplification, protein isolation for malaria detection, cell counting for human immunodeficiency virus diagnosis, and clinical chemistry assays. 21,22,[27][28][29] In each of these assays, a single magnet was moved relative to the length of the tube to transfer the magnetic beads across the immiscible fluid interface and into the sequential processing chambers. We have recently described a twomagnet design for processing which worked particularly well for viscous sputum samples used for tuberculosis diagnosis.…”
Section: Introductionmentioning
confidence: 99%
“…The depositions tube’s millimeter-diameter tube aperture creates a stable surface tension valve that retained the sample and allowed beads to collect at the meniscus formed at the bottom of the tube during the 1 min bead collection. 32 The developed surface tension valve prevents premature transfer of magnetic beads to the LFA during magnetic bead gathering step. The cantilever-based tube holder was then depressed briefly by the user in order for the bead-filled meniscus to make contact with the conjugate pad, breaking the surface tension valve, and transferring the beads to the LFA.…”
Section: B Experimental Methodsmentioning
confidence: 99%
“…Thus, the total measurement time with the chip is greatly reduced, totaling approximately 40 min, which is much shorter than the time needed to run a conventional plate-based ELISA (more than 4 h). In another bead-based microfluidic assay approach using surface tension valves for fluidic control, 33 the measurement of the concentration of the biomarker in that chip was performed off-chip, but in the proposed chip, it was performed in the same chip, which eliminates the need for manual operation and facilitates automatic control of the measurement process. 96-well plate.…”
Section: B Detection Of the Bladder Cancer Biomarker Apoa1mentioning
confidence: 99%