CR-39 detector-based Radon dosimetry system calibration in the self-decay mode

Abstract: This paper describes cooperation of the Slovak University of Technology in Bratislava (STU) and the Slovak Institute of Metrology (SMU) in the verification of the calibration factors for the solid-state nuclear track detector utilized for radon dosimetry and establishing traceability for the STU laboratory. The SMU operates the Air Radon Standard which consists of a radon chamber and calibrated radon atmosphere monitoring system. A specification of the radon chamber is that the radon concentration changes duri… Show more