Background-The upper limit of vulnerability (ULV) correlates with the defibrillation threshold and can be determined with 1 episode of ventricular fibrillation (VF). To automate the ULV in an implantable cardioverter-defibrillator (ICD), the most vulnerable intervals must be identified from an ICD electrogram rather than the latest-peaking surface T wave (T peak ). We hypothesized that the recovery time (T R ), defined as the maximum derivative (dV/dt) of the T wave of the shock electrogram, correlates with the most vulnerable intervals. Methods and Results-We determined ULV, defibrillation threshold, and the most vulnerable intervals in 25 patients at ICD implantation. The ULV was the weakest T-wave shock that did not induce VF. The most vulnerable intervals were the ones associated with the strongest shocks that induced VF. Telemetered shock electrograms were stored on digital tape and differentiated offline to measure T R . T peak and T R were highly correlated (T peak ϪT R ϭϪ2Ϯ11 ms; ϭ0.80, PϽ0.001). At least 1 most vulnerable interval timed between Ϫ20 ms and ϩ20 ms relative to T peak in all patients and between Ϫ40 ms and ϩ20 ms relative to T R in 96% of patients.