2019
DOI: 10.3390/diagnostics9030094
|View full text |Cite
|
Sign up to set email alerts
|

Comparison of Procalcitonin Assays on KRYPTOR and LIAISON® XL Analyzers

Abstract: Our laboratory performs procalcitonin (PCT) assays on a Brahms KRYPTOR analyzer with the Brahms PCT sensitive Kryptor kit. In this study, we wanted to compare the assays obtained in this way with the ones performed on the LIAISON® XL. From January to May 2017, 171 samples were analyzed, of which 65 from female patients (age: 22–98 years) and 106 from male patients (age: 16–97 years). The PCT determination was performed using the LIAISON® XL and KRYPTOR analyzers, by chemiluminescence (Chemiluminescence immunoa… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0
1

Year Published

2020
2020
2023
2023

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 11 publications
(6 citation statements)
references
References 7 publications
0
5
0
1
Order By: Relevance
“…İnsan kanındaki PCT konsantrasyonunu saptamak için, çoğunlukla kemilüminesans (21), immünofloresan (22), immünoluminometri (23) ve ayrıca yüzey plazmon rezonans biyosensörlerini (24) kullanan çeşitli yöntemler mevcuttur. Bu yöntemlerin tümü, aynı monoklonal fare anti-katakalsin antikorlarının ve monoklonal veya poliklonal koyun anti-kalsitonin antikorlarının (tahliller arasında farklıdır) bir kombinasyonuna dayanır.…”
Section: Prokalsi̇toni̇nunclassified
“…İnsan kanındaki PCT konsantrasyonunu saptamak için, çoğunlukla kemilüminesans (21), immünofloresan (22), immünoluminometri (23) ve ayrıca yüzey plazmon rezonans biyosensörlerini (24) kullanan çeşitli yöntemler mevcuttur. Bu yöntemlerin tümü, aynı monoklonal fare anti-katakalsin antikorlarının ve monoklonal veya poliklonal koyun anti-kalsitonin antikorlarının (tahliller arasında farklıdır) bir kombinasyonuna dayanır.…”
Section: Prokalsi̇toni̇nunclassified
“…However, not all assays commercially available take part in this harmonization effort, thus there is currently no commonly agreed reference material for all available PCT immunoassays at the international level. Therefore, the need for a new reference system based on higher order reference materials and/or a higher-order reference method to achieve global standardization/harmonization of PCT results worldwide has been pointed out in a number of studies [56][57][58][59]64,65].…”
Section: Procalcitonin and Its Current Measurement Proceduresmentioning
confidence: 99%
“…Short detection time with 30 min is the biggest advantage. 39) Recently, quantitative assay test by automated immunofluorescence measuring instrument based on the chemiluminescence immunoassay method is available. 40) PCT attracts a lot of attention as a specific bacterial infection marker, 37) but sensitivity and specificity in Sepsis is not so significant compare to C-reactive protein (CRP).…”
Section: Procalcitonin Rapid Test Reagentmentioning
confidence: 99%