Hepcidin concentrations measured by various methods differ considerably, complicating interpretation. Here, a previously identified plasma-based candidate secondary reference material (csRM) was modified into a serum-based two-leveled sRM. We validated its functionality to increase the equivalence between methods for international standardization.
Methods
We applied technical procedures developed by the International Consortium for Harmonization of Clinical Laboratory Results. The sRM, consisting of lyophilized serum with cryolyoprotectant, appeared commutable among nine different measurement procedures using 16 native human serum samples in a first round robin (RR1). Harmonization potential of the sRM was simulated in RR1 and evaluated in practice in RR2 among 11 measurement procedures using three native human plasma samples. Comprehensive purity analysis of a candidate primary RM (cpRM) was performed by state of the art procedures. The sRM was value assigned with an isotope dilution mass spectrometry-based candidate reference method calibrated using the certified pRM.
Results
The inter-assay CV without harmonization was 42.1% and 52.8% in RR1 and RR2, respectively. In RR1, simulation of harmonization with sRM resulted in an inter-assay CV of 11.0%, whereas in RR2 calibration with the material resulted in an inter-assay CV of 19.1%. Both the sRM and pRM passed international homogeneity criteria and showed long-term stability. We assigned values to the low (0.95±0.11 nmol/L) and middle concentration (3.75±0.17 nmol/L) calibrators of the sRM.
Conclusions
Standardization of hepcidin is possible with our sRM, which value is assigned by a pRM. We propose the implementation of this material as an international calibrator for hepcidin.
Chronic kidney disease definition is based on glomerular filtration rate (GFR) estimations which are derived from creatinine-based equations. The accuracy of GFR estimation is thus largely dependent of those of serum creatinine assays. International recommendations highlight the need for traceable creatinine assays. The French Society of Clinical Biochemistry conducted a study for measuring accuracy of creatinine enzymatic methods. This evaluation involved 25 clinical laboratories. Creatinine was measured in serum pools ranging from 35.9 ± 0.9 μmol/L to 174.5 ± 3.1 μmol/L (IDMS determination) using 12 creatinine enzymatic methods. For all creatinine values greater than 74.4 ± 1.4 μmol/L, the bias and imprecision did not exceed 5% and 5.9%, respectively. For the lowest value (35.9 ± 0.9 μmol/L), the bias ranged from −1.8 to 9.9% (with one exception). At this level, the imprecision ranged from 1.9 to 7.8%. The true performances of the assays (couples of bias and relative standard deviation), were evaluated using Monte-Carlo simulations. Most of the assays fall within the maximum Total Error of 12% at all concentrations. This study demonstrates substantial improvements in the calibration, traceability and precision of the enzymatic methods, reaching the NKDEP recommendations. Moreover, most of these assays allowed accurate creatinine measurements for creatinine levels lower than 40 μmol/L.
Commutability is a property of a reference material (RM) that relates to the closeness of agreement between results for an RM and results for clinical samples (CSs) when measured by ≥ 2 measurement procedures (MPs). Commutability of RMs used in a calibration traceability scheme is an essential property for them to be fit for purpose. Similarly, commutability of trueness controls or external quality assessment samples is essential when those materials are used to assess trueness of results for CSs. This report is part 1 of a 3-part series describing how to assess commutability of RMs. Part 1 defines commutability and addresses critical components of the experimental design for commutability assessment, including selection of individual CSs, use of pooled CSs, qualification of MPs for inclusion, establishing criteria for the determination that an RM is commutable, generalization of commutability conclusions to future measurements made with the MPs included in the assessment, and information regarding commutability to be included in the certificate for an RM. Parts 2 and 3 in the series present 2 different statistical approaches to commutability assessment that use fixed criteria related to the medical decisions that will be made using the laboratory test results.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.