2017
DOI: 10.1016/j.apradiso.2016.12.036
|View full text |Cite
|
Sign up to set email alerts
|

Compact radioactive aerosol monitoring device for early warning networks

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2

Citation Types

1
3
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 4 publications
1
3
0
Order By: Relevance
“…The values obtained by Monte Carlo simulation of the experimental measurements described in [6] are 1.36E−03 for 661.6 keV and 7.21E−04 for 1115.5 keV. These deviate from the experimental values for −2.08% and −2.1% respectively which may be considered a satisfactory match that validates the simulation.…”
Section: Resultssupporting
confidence: 58%
See 3 more Smart Citations
“…The values obtained by Monte Carlo simulation of the experimental measurements described in [6] are 1.36E−03 for 661.6 keV and 7.21E−04 for 1115.5 keV. These deviate from the experimental values for −2.08% and −2.1% respectively which may be considered a satisfactory match that validates the simulation.…”
Section: Resultssupporting
confidence: 58%
“…For 661.6 keV (Cs-137) and 1115.5 keV (Zn-65) photons the efficiencies reported in the paper [6] are 1.38E−03 and 7.36E−04 counts per photon emitted.…”
Section: Resultsmentioning
confidence: 78%
See 2 more Smart Citations