Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

3
102
5
4

Year Published

2003
2003
2021
2021

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 157 publications
(114 citation statements)
references
References 39 publications
3
102
5
4
Order By: Relevance
“…Along the leaf end these values were 2.22 and 2.41 mm for the 9.1×9.1 and 8.0×8.0cm2 fields, respectively. These values are slightly smaller than those reported by Cosgrove et al (1) (2.45 mm perpendicular to leaf motion and 2.60 mm parallel with leaf motion) for square fields when using radiographic film X‐OMAT V2. This might be attributed to the energy dependence of the radiographic film which produces broadening of penumbra measurements (7) …”
Section: Discussioncontrasting
confidence: 72%
See 1 more Smart Citation
“…Along the leaf end these values were 2.22 and 2.41 mm for the 9.1×9.1 and 8.0×8.0cm2 fields, respectively. These values are slightly smaller than those reported by Cosgrove et al (1) (2.45 mm perpendicular to leaf motion and 2.60 mm parallel with leaf motion) for square fields when using radiographic film X‐OMAT V2. This might be attributed to the energy dependence of the radiographic film which produces broadening of penumbra measurements (7) …”
Section: Discussioncontrasting
confidence: 72%
“…The use of several detectors, such as small ionization chambers and radiographic films (X‐OMAT V2 and EDR2) for the characterization of basic dosimetric properties of mMLC systems, such as leaf transmission, leakage and beam penumbra measurements, has been reported extensively in the literature 1 , 6 . Small ionization chambers provide high precision measurements for field sizes larger than their effective active area.…”
Section: Introductionmentioning
confidence: 99%
“…1 should be equal to arctan (5/100), or 2.9°. This contradicts the 1.8° value presented by Belec et al ( 21 ) but compares favorably to approximately 3° quoted by Cosgrove et al ( 1 ) …”
Section: Methodscontrasting
confidence: 91%
“…1). Cosgrove et al ( 1 ) explained how the leaves are focused towards the target; the central ray is parallel to the vertical edge with the leaf at isocenter, and to the slanted edges when the leaf is extended to 5cm or retracted to +5cm. For the edges to line up with the central ray under these conditions, the slant angle Θ in Fig.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation