2014
DOI: 10.14319/ijcto.0204.15
|View full text |Cite
|
Sign up to set email alerts
|

Commissioning and quality assurances of the Intrabeam Intra-Operative radiotherapy unit

Abstract: Purpose: The authors report comprehensive commissioning and quality assurance (QA) procedures for Intrabeam, Intra-Operative radiotherapy (IORT) unit. The Intrabeam system miniature X-ray source is a 50 kV and 40 µA unit. Methods: The authors' tests include measurements of output, beam deflection, isotropy, kVp and mAs measurements, quality index, isodose, reproducibility, linearity, depth dose verification, and 3D dose distribution. IC ionization chamber and the UNIDOSE dosimeter were used for the output comm… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
5
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(5 citation statements)
references
References 10 publications
0
5
0
Order By: Relevance
“…The latter include comprehensive testing and quality assurance of all delivery equipment along with beam data acquisition and validation. 41 The commissioning process may be divided into two broad categories: mechanical and radiation checks. Mechanical tests include the integrity checks of cables, QA devices, treatment applicators, robotic x-ray stand (including stand balancing and precision positioning tests), and software related to QA tests.…”
Section: System Commissioningmentioning
confidence: 99%
See 3 more Smart Citations
“…The latter include comprehensive testing and quality assurance of all delivery equipment along with beam data acquisition and validation. 41 The commissioning process may be divided into two broad categories: mechanical and radiation checks. Mechanical tests include the integrity checks of cables, QA devices, treatment applicators, robotic x-ray stand (including stand balancing and precision positioning tests), and software related to QA tests.…”
Section: System Commissioningmentioning
confidence: 99%
“…The radiation protection survey is followed by IORT commissioning measurements. The latter include comprehensive testing and quality assurance of all delivery equipment along with beam data acquisition and validation 41 …”
Section: System Commissioningmentioning
confidence: 99%
See 2 more Smart Citations
“…Through these developments, electronic brachytherapy system has emerged as one of the most rapidly disseminating medical devices in the world. However, the core component of these systems is still the miniature tube that allows the low energy x-ray to be produced at high dose rates ( 20 ). The x-ray spectrum produced by this tube has a continuous distribution (bremsstrahlung effect) with a line spectrum characteristic ( 21 ), and has an extensive use in medical physics ( 22 ).…”
Section: Introductionmentioning
confidence: 99%