2024
DOI: 10.1002/acm2.14272
|View full text |Cite
|
Sign up to set email alerts
|

Intraoperative radiation therapy with 50 kV x‐rays: A multi‐institutional review

Anil Sethi,
Sebastien Gros,
Patrik Brodin
et al.

Abstract: This report covers clinical implementation of a low kV intraoperative radiation therapy (IORT) program with the INTRABEAM® System (Carl Zeiss Meditec AG, Jena, Germany). Based on collective user experience from eight institutions, we discuss best methods of INTRABEAM quality assurance (QA) tests, commissioning measurements, clinical workflow, treatment planning, and potential avenues for research. The guide provides pertinent background information and clinical justification for IORT. It describes the INTRABEA… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
0
0

Publication Types

Select...

Relationship

0
0

Authors

Journals

citations
Cited by 0 publications
references
References 82 publications
(166 reference statements)
0
0
0
Order By: Relevance

No citations

Set email alert for when this publication receives citations?