Clinical evaluation of the Canon TX-10 noncontact tonometer in healthy eyes

Abstract: There was a good agreement between Canon TX-10 and GAT in healthy subjects. From the results obtained here, CCT appears to not affect IOP measurements when CCT is thinner than 570 microm.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4

Citation Types

6
18
0

Year Published

2010
2010
2023
2023

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 19 publications
(24 citation statements)
references
References 41 publications
6
18
0
Order By: Relevance
“…Clinical agreement between the noncontact tonometer and the GAT has been demonstrated in previous studies of normotensive subjects,5,6,29,30 with a propensity for the noncontact tonometer to return about 3 mmHg higher IOP measurements than the GAT7,8,27,28 in subjects with IOPs up to 21 mmHg.…”
Section: Discussionsupporting
confidence: 59%
See 1 more Smart Citation
“…Clinical agreement between the noncontact tonometer and the GAT has been demonstrated in previous studies of normotensive subjects,5,6,29,30 with a propensity for the noncontact tonometer to return about 3 mmHg higher IOP measurements than the GAT7,8,27,28 in subjects with IOPs up to 21 mmHg.…”
Section: Discussionsupporting
confidence: 59%
“…Overall, both instruments showed good repeatability, with the PT100 noncontact tonometer being more consistent in repeated measurement sessions. Two other studies5,6 have also shown similar repeatability coefficients for the GAT, ±2.5 mmHg 5 and 3.70 mmHg 6 when compared with other noncontact tonometers 5,6…”
Section: Discussionsupporting
confidence: 56%
“…Studies on normal eyes have demonstrated CoR values ranging from 2.2 mmHg to 9.2 mmHg and 95% limits of agreement ranging from −7.6 mmHg to +14.2 mmHg between various NCTs and GAT [10,[18][19][20][21][22][23]. In general, NCTs tended to over-read the GAT IOP measurements.…”
Section: Discussionmentioning
confidence: 91%
“…Several non-contact tonometers (NCTs) for the IOP assessment have been developed, showing variable statistics on the reliability of the various techniques [10,[17][18][19][20][21][22][23].…”
Section: Discussionmentioning
confidence: 99%
“…9 The previous generation of the Reichert (Reichert Xpert and Reichert AT550) NCT demonstrated good agreement with GAT in healthy and glaucoma patients. [10][11][12][13][14][15] The R7 uses the same formula as the ORA for measuring the IOP according to the manufacturer.…”