Chlormequat Selective Electrodes: Construction, Evaluation and Application at Fia Systems

Abstract: A flow injection analysis (FIA) system having a chlormequat selective electrode is proposed. Several electrodes with poly(vinyl chloride) based membranes were constructed for this purpose. Comparative character-ization suggested the use of membrane with chlormequat tetraphenylborate and dibutylphthalate. On a single-line FIA set-up, operating with 1 x 10 -2 mol L -1 ionic strength and 6.3 pH, calibration curves presented slopes of 53.6 ± 0.4 mV decade -1 within 5.0 x 10 -6 and 1.0 x 10 -3 mol L -1, and squared… Show more