Calibration of Digital Thermometer using Sensor and Indicator Method

Abstract: A Calibration has been done for digital thermometer using a sensor and indicator method. The unit under test manufactured by Fluke with 54-II type and serial number 13160058, meanwhile, the standards that used are PT100 sensor from Omega and an indicator from Additel. The calibration result from 12 setpoints shows the average correction value as 0.61 ºC with uncertainty as 0.32 ºC and K factor as 2.0 at 95% confidence level

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
0
0
1

Year Published

2023
2023
2023
2023

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(1 citation statement)
references
References 2 publications
(2 reference statements)
0
0
0
1
Order By: Relevance
“…Data logger yang digunakan telah dilakukan kalibrasi minimal 1 tahun sekali. Kalibrasi dilakukan dengan tujuan untuk menentukan nilai yang sebenarnya dari alat ukur yang dibandingkan dengan standar alat ukur baik standar nasional maupun internasional [27,28]. Variasi suhu hasil rekaman data logger akan disimpan dan dianalisis menggunakan komputer.…”
Section: Pembahasanunclassified
“…Data logger yang digunakan telah dilakukan kalibrasi minimal 1 tahun sekali. Kalibrasi dilakukan dengan tujuan untuk menentukan nilai yang sebenarnya dari alat ukur yang dibandingkan dengan standar alat ukur baik standar nasional maupun internasional [27,28]. Variasi suhu hasil rekaman data logger akan disimpan dan dianalisis menggunakan komputer.…”
Section: Pembahasanunclassified