“…Results obtained with digestion performed with the proposed procedure presented a good agreement with certified values (higher than 96%), even if digestion was performed with a solution of 2 mol L À1 HNO 3 . As observed in previous works [19][20][21] the use of diluted nitric acid and vessels pressurized with oxygen resulted in low blank values. For 500 mg and using 2 mol L À1 HNO 3 and 0.5 MPa O 2 , low blank values were obtained allowing limits of detection of 12, 18, 18, 12, 10, and 12 ng g À1 for Ca, Cu, Fe, Mg, Mn, and Zn, respectively.…”