2021
DOI: 10.52711/0974-360x.2021.00613
|View full text |Cite
|
Sign up to set email alerts
|

Analytical Method Development and Validation of Etravirine by RP-UFLC

Abstract: An ultra-fast liquid chromatography (RP-UFLC) approach was processed for the estimation of ETR, which is quick, responsive, reliable, and inexpensive. Work on the Phenomenex Kinetex C18 column (250x 4,6mm. 5μ) was carried out with MeOH and acetonitrile in the mobile phase ratio of (60:40v/v) at 1.0mL/min. With a PDA detection system, the eluent was tracked at 311nm. ETR elutes at a persistence time of 3.226 min. The proposed method gives linearity of concentration from 1 to 5μg/mL, with the value of R2 at 0.99… Show more