2020
DOI: 10.1088/1361-6560/ab8534
|View full text |Cite
|
Sign up to set email alerts
|

An integrated quality assurance phantom for frameless single-isocenter multitarget stereotactic radiosurgery

Abstract: Brain stereotactic radiosurgery (SRS) treatments require multiple quality assurance (QA) procedures to ensure accurate and precise treatment delivery. As single-isocenter multitarget SRS treatments become more popular, the quantification of off-axis accuracy of the linear accelerator is crucial. In this study, a novel brain SRS integrated phantom was developed and validated to enable SRS QA with a single phantom to facilitate implementation of a frameless single-isocenter, multitarget SRS program. This phantom… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
13
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 17 publications
(13 citation statements)
references
References 42 publications
0
13
0
Order By: Relevance
“…15 In many cases, these phantoms cannot be irradiated from certain angles since the electronics are sensitive to radiation. 15,19,22 This limits their ability to measure true composite plans for patient-specific QA as recommended by the AAPM report TG-218. These phantoms also typically suffer from artifacts when scanned with a CT scanner so they must be virtually modeled in the treatment planning system until the results agree with measurements.…”
Section: C Electronic Phantomsmentioning
confidence: 99%
See 3 more Smart Citations
“…15 In many cases, these phantoms cannot be irradiated from certain angles since the electronics are sensitive to radiation. 15,19,22 This limits their ability to measure true composite plans for patient-specific QA as recommended by the AAPM report TG-218. These phantoms also typically suffer from artifacts when scanned with a CT scanner so they must be virtually modeled in the treatment planning system until the results agree with measurements.…”
Section: C Electronic Phantomsmentioning
confidence: 99%
“…Various attempts have been made utilizing different commercial and non-commercial system to varying degrees of success. [10][11][12][13][14][15][16][17][18][19][20][21][22] Some of the more common and current methods and their limitations will be briefly discussed (Figs. 1-7).…”
Section: Introductionmentioning
confidence: 99%
See 2 more Smart Citations
“…However, some QA programs are tedious and timeconsuming to complete with repeated phantom displacements and manual measurements. To enhance the QA efficiency for machine performance, streamlined QA is clinically essential and various methods have been proposed (7)(8)(9)(10). For instance, Varian has released a commercial module named Machine Performance Check (MPC) (Varian, CA, USA) for its high-end LINAC systems.…”
Section: Introductionmentioning