An experimental setup for traceable measurement and calibration of liquid flow rates down to 5 nl/min

Abstract: This work presents the improvements of an experimental setup for measuring ultra-low flow rates down to 5 nl/min. The system uses a telecentric CCD imaging system mounted on a high-precision, computer-controlled linear stage to track a moving liquid meniscus inside a glass capillary. Compared to the original setup, the lowest attainable expanded uncertainty at any flow rate has been reduced from 5.4% to 2%. In addition, the conformity with specification of three commercial micro-fluidic devices was evaluated u… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
8
0

Year Published

2015
2015
2024
2024

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 9 publications
(9 citation statements)
references
References 4 publications
0
8
0
Order By: Relevance
“…In response to the limitations of the gravimetric principle previously mentioned, investigations on the front tracking meniscus method were carried on by Lubeck University. 110,113 First investigated for low flow rate by Richter et al, 114 this method relies on tracking the meniscus displacement of a flowing liquid in a capillary of known dimensions set at the outlet of the calibration system, as described in Fig. 7.…”
Section: Front Tracking Meniscus (Ftm)mentioning
confidence: 99%
See 2 more Smart Citations
“…In response to the limitations of the gravimetric principle previously mentioned, investigations on the front tracking meniscus method were carried on by Lubeck University. 110,113 First investigated for low flow rate by Richter et al, 114 this method relies on tracking the meniscus displacement of a flowing liquid in a capillary of known dimensions set at the outlet of the calibration system, as described in Fig. 7.…”
Section: Front Tracking Meniscus (Ftm)mentioning
confidence: 99%
“…On the left: Side view. 113 measurements and standardization. This is addressed by metrology, but there is still a delay and a performance gap between the development of flowmeters by academic and industrial players, and the development of primary standards within metrological institutes.…”
Section: Perspectivesmentioning
confidence: 99%
See 1 more Smart Citation
“…The position of the pump must be adjusted to the height of the meniscus of the oil layer in the glass vial [2,8], which is required by the standard IEC 60601-2-24:2012 [4]. The optical method is the front track of the meniscus of a liquid in a pipette or a capillary, using the position of the meniscus as a function of time to determine the displacement and its corresponding volume of one or several consecutive doses [6,7,9,10]. Another optical method is to determine the volume of a liquid sphere attached to the cannula.…”
Section: Introductionmentioning
confidence: 99%
“…In addition to the common gravimetric method, new methods are discussed that could be applied for extremely low flows in case the gravimetric method is not suitable. This includes an optical front tracking method [1] and flow sources. One source based on volume expansion is introduced in [9].…”
mentioning
confidence: 99%