1998
DOI: 10.1007/s002709900216
|View full text |Cite
|
Sign up to set email alerts
|

Amplatz Thrombectomy Device: Unexpected Interaction with the Cragg Endoprosthesis

Abstract: In a 74-year-old woman, a Dacron-covered Cragg endoprosthesis was implanted into the left superficial femoral artery after successful recanalization of a 6-cm-long occlusion. At 4.5 months after discharge the endoprosthesis was occluded. In order to perform a thrombectomy, an Amplatz thrombectomy device (ATD) was activated and advanced into the occluded graft. Suddenly, a mechanical failure of the ATD occurred. Extraction of the ATD was achieved only after repeated attempts at rotating motion and jerky pullbac… Show more