2001
DOI: 10.3109/10929080109146093
|View full text |Cite
|
Sign up to set email alerts
|

Accuracy in Tunnel Placement for ACL Reconstruction. Comparison of Traditional Arthroscopic and Computer-Assisted Navigation Techniques

Abstract: The purpose of this randomized, prospective study was to compare accuracy in tunnel placement as performed with a traditional arthroscopic anterior cruciate ligament (ACL) reconstruction technique and with KneeNavTM ACL, a computer-assisted surgical navigation technique. Two surgeons experienced in ACL reconstruction, but inexperienced in computer-assisted surgical navigation technique, each randomly used traditional arthroscopic guides or KneeNavTM ACL to drill a tunnel in twenty identical foam knees. Placeme… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

3
50
0

Year Published

2006
2006
2018
2018

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 111 publications
(53 citation statements)
references
References 12 publications
3
50
0
Order By: Relevance
“…These results were comparable with those from several other studies which have quantified the accuracy or precision of ACL navigation systems [40,[43][44][45]. However, the measurements in three of those studies only compared the position of the preoperatively planned tunnels with the postoperative results, not with the actual position of the anatomical insertion centers; two of the studies [43,44] used sawbones, and one [45] used cadaveric knees. Musahl et al [42] defined the accuracy and precision of preoperative tunnel planning with an active robotic system in relation to a tunnel placed in the actual femoral ACL insertion, the values being 1.3 mm (mean) and 1.0 (SD), respectively.…”
Section: Discussionsupporting
confidence: 71%
See 3 more Smart Citations
“…These results were comparable with those from several other studies which have quantified the accuracy or precision of ACL navigation systems [40,[43][44][45]. However, the measurements in three of those studies only compared the position of the preoperatively planned tunnels with the postoperative results, not with the actual position of the anatomical insertion centers; two of the studies [43,44] used sawbones, and one [45] used cadaveric knees. Musahl et al [42] defined the accuracy and precision of preoperative tunnel planning with an active robotic system in relation to a tunnel placed in the actual femoral ACL insertion, the values being 1.3 mm (mean) and 1.0 (SD), respectively.…”
Section: Discussionsupporting
confidence: 71%
“…The reported precision is similar to our own result, while the accuracy seems superior. However, the system also showed a deviation of 1.3 mm between preoperatively planned tunnels and postoperative results [45].…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…In the published orthopaedic literature, computer-assisted surgery has been reported to raise the accuracy of tunnel placement and reduce the rate of revision surgery. Picard et al [21] presented a randomised prospective study comparing the accuracy of tunnel placement performed using a traditional arthroscopic ACL reconstruction technique and a surgical navigation technique. Two experienced surgeons in ACL reconstruction randomly used traditional arthroscopic guides or navigated assisted guides to drill a tunnel in 20 identical foam knees.…”
Section: Discussionmentioning
confidence: 99%