A synthetic diamond diode in volumetric modulated arc therapy dosimetry

Abstract: Purpose: The aim of this work is to investigate the behavior of a single crystal diamond diode (SCDD) for volumetric modulated arc therapy (VMAT) dose verifications. This delivery technique is one of the most severe test of a dosimeter performance due to the modulation of the dose rate achieved by simultaneously changing the velocity of the gantry and the position of the collimator leaves. The performed measurements with VMAT photon beams can therefore contribute to an overall global validation of the device t… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
14
0
1

Year Published

2014
2014
2020
2020

Publication Types

Select...
9

Relationship

1
8

Authors

Journals

citations
Cited by 21 publications
(15 citation statements)
references
References 28 publications
(29 reference statements)
0
14
0
1
Order By: Relevance
“…The use of research SCDDs has been reported [6][7][8] in conjunction with conventional linear accelerators, with square field sizes from 100 3 100 mm 2 down to 10 3 10 mm 2 as a measured at a 1000-mm source-to-detector distance (SDD). In this work, we present the results using a dedicated stereotactic linear accelerator with circular fields for fixed collimators ranging from 5 to 60 mm, at a nominal SDD of 800 mm, in a 6-MV unflattened photon beam.…”
Section: Methodsmentioning
confidence: 99%
“…The use of research SCDDs has been reported [6][7][8] in conjunction with conventional linear accelerators, with square field sizes from 100 3 100 mm 2 down to 10 3 10 mm 2 as a measured at a 1000-mm source-to-detector distance (SDD). In this work, we present the results using a dedicated stereotactic linear accelerator with circular fields for fixed collimators ranging from 5 to 60 mm, at a nominal SDD of 800 mm, in a 6-MV unflattened photon beam.…”
Section: Methodsmentioning
confidence: 99%
“…No voltage bias is required and the diamond operates as a Schottky diode. [30][31][32][33][34] The set of measurements consisted in point data acquisition, on the beam central axis at isocenter at 10 cm depth (source to surface distance SSD = 90 cm). Precise positioning of the detector on the beam central axis was guaranteed by centering the system according to the profiles acquired along the main axes at the measuring depth.…”
Section: A Experimental Measurementsmentioning
confidence: 99%
“…[15][16][17][18] The microDiamond active volume is composed of a 1 µm thick intrinsic diamond layer, with a 1 mm radius and a total sensitive volume of 0.004 mm 3 . The effective point of measurement is 1 mm from the housing surface.…”
Section: A1 Microdiamond Ptw-60019mentioning
confidence: 99%
“…The characteristics of the synthetic diamond detector (PTW-60019), such as small active volume (0.004 mm 3 ), near water-equivalence (Z eff = 6), and low energy dependence, make it an interesting candidate for small field dose measurements as demonstrated by many authors. [15][16][17][18][19] To our knowledge, no one demonstrated the feasibility of using this synthetic diamond for OFs measurement for LGKP. Profiles in the three directions were also measured for the 4 mm collimator.…”
Section: Introductionmentioning
confidence: 99%