2015
DOI: 10.1007/s13738-015-0756-z
|View full text |Cite
|
Sign up to set email alerts
|

A soft-template nanostructured peroxidase based on cytochrome c and sodium decyl sulfate and its electrochemical properties on hydroxyl fullerenes modified glassy carbon electrode

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
4
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
5
2

Relationship

3
4

Authors

Journals

citations
Cited by 7 publications
(5 citation statements)
references
References 52 publications
1
4
0
Order By: Relevance
“…The regression equations were I pc = 0.036 ν + 1.243 (R 2 = 0.9965) and I pa = −0.036 ν − 1.668 (R 2 = 0.995). In Figure 5 C, the absolute slopes value of log I vs. log ν were close to 1, so these indicate that the electrochemical behavior of rutin on the electrode surface is an adsorption-controlled process [ 16 ], which is the same as in other reports [ 51 , 52 ]. This phenomenon suggests that the determination of rutin should start from a low concentration, or the adsorbed rutin in the blank phosphate-buffered solution should be removed after detecting a high concentration of rutin.…”
Section: Resultssupporting
confidence: 81%
“…The regression equations were I pc = 0.036 ν + 1.243 (R 2 = 0.9965) and I pa = −0.036 ν − 1.668 (R 2 = 0.995). In Figure 5 C, the absolute slopes value of log I vs. log ν were close to 1, so these indicate that the electrochemical behavior of rutin on the electrode surface is an adsorption-controlled process [ 16 ], which is the same as in other reports [ 51 , 52 ]. This phenomenon suggests that the determination of rutin should start from a low concentration, or the adsorbed rutin in the blank phosphate-buffered solution should be removed after detecting a high concentration of rutin.…”
Section: Resultssupporting
confidence: 81%
“…6). The detection limit based on 3s b (s b was the standard deviation of ten replicate measurements of fabricated biosensor in [Fe(CN) 6 3−/4− ] solution) was obtained to be 2.5 × 10 −15 M. These results shows that GO due to their unique properties such as having lots of carboxylic acid groups, increasing the surface area, good biocompatibility and electron mobility at room temperature exhibits high sensitivity. Also, some characteristics of ssDNA/GO/GCE biosensor were compared with the reported electrodes in Table 1.…”
Section: Investigation Of Sensitivity Of Modified Electrodementioning
confidence: 78%
“…From the slope of the I peak versus ν 1/2 at various scan 6 ] solution, the microscopic areas for GCE and GO/GCE were evaluated to be 0.0701 and 0.0910 cm 2 , respectively. Also, the current densities of modified and non-modified electrodes were calculated.…”
Section: Characterization Of Modified Electrodementioning
confidence: 99%
“…It could be seen that the linear range for measurement of GOD activity was from 4.3×10 -5 mM to 9.3×10 -3 mM (Fig.5A). The detection limit was 4.3×10 -5 mM (Fig.5B) [20][21][22]. …”
Section: E Linear Range and Detection Limitmentioning
confidence: 99%