A pulsed magnetomotive ultrasound imaging system for magnetic nanoparticle detection

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1
1

Relationship

1
1

Authors

Journals

citations
Cited by 2 publications
(7 citation statements)
references
References 10 publications
0
7
0
Order By: Relevance
“…After charging the capacitor, an electronic switching device and the coil generated the magnetic field pulse. For further information about the MMUS setup, please refer to Mazon et al [ 44 ]. A multichannel ultrasound pulse/echo system (Sonix RP + Sonix DAQ, Ultrasonix) was then used to track the induced displacement of the internal structure (in the order of micrometers) by a cross-correlation method [ 45 ].…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…After charging the capacitor, an electronic switching device and the coil generated the magnetic field pulse. For further information about the MMUS setup, please refer to Mazon et al [ 44 ]. A multichannel ultrasound pulse/echo system (Sonix RP + Sonix DAQ, Ultrasonix) was then used to track the induced displacement of the internal structure (in the order of micrometers) by a cross-correlation method [ 45 ].…”
Section: Methodsmentioning
confidence: 99%
“…It should be mentioned that the US acquisition was synchronized with the magnetic excitation through a computer using a LabVIEW interface. This system operated with a frame rate of 4 kHz, and the magnetic pulse duration varied from 4 to 8 ms [ 44 ]. The maximum magnetic field applied 2 mm from the tip of the core was 740 mT.…”
Section: Methodsmentioning
confidence: 99%
“…Nevertheless, their implementation involves high costs and special infrastructure installation and still presents minimum load impedance defined by the output power. Recently, other studies using pulsed magnetic field generators based on capacitor discharge [25,52,53] have shown improvements in magnetic field features such as field magnitude up to 950 mT, acceptable variable pulse duration (2 ms -1s), and capabilities of driving low-impedance coils, lower than 2 Ω.…”
Section: Chapter 2: Development Of Pulsed Magnetomotive Ultrasound Im...mentioning
confidence: 99%
“…Therefore, by increasing the magnetic field generated, the induced displacement is enhanced and, consequently, improving the MMUS imaging contrast and the signal-to-noise ratio [26,52]. Another important term is magnetic field time dependence, which is associated with the magnetic field waveform and attractive magnetic force.…”
Section: A Magnetic Forcementioning
confidence: 99%