A novel portable energy dispersive X-ray fluorescence spectrometer with triaxial geometry

Abstract: The X-ray fluorescence technique is a powerful analytical tool with a broad range of applications such as quality control, environmental contamination by heavy metals, cultural heritage, among others. For the first time, a portable energy dispersive X-ray fluorescence spectrometer was assembled, with orthogonal triaxial geometry between the X-ray tube, the secondary target, the sample and the detector. This geometry reduces the background of the measured spectra by reducing significantly the Bremsstrahlung pro… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
9
0

Year Published

2017
2017
2022
2022

Publication Types

Select...
5
2

Relationship

2
5

Authors

Journals

citations
Cited by 21 publications
(9 citation statements)
references
References 29 publications
0
9
0
Order By: Relevance
“…An experimental setup was assembled using a sided window X-ray tube with a Mo anode, OXFORD XTF5011 (max 50 kV, 1 mA) and a Silicon Drift Detector, thermoelectrically cooled (60EX, Vortex) (FWHM at 160 eV at Fe-Kα), with an active area of 50 mm 2 and a 25 μm thickness Be window. Monochromatization and reduction of the Bremstrahlung from the tube was obtained using a secondary Yttrium target disk (99% purity, 1.0 mm thickness, 25.0 mm diameter, GoodFellow Inc., Huntingdon, UK) in triaxial geometry as it was described in a previous work (Pessanha et al, 2017). The spot size was measured with a radiographic film producing an ellipsoid image with 17 mm × 12 mm.…”
Section: 2portable Setup With Triaxial Geometry -P3xrfmentioning
confidence: 99%
“…An experimental setup was assembled using a sided window X-ray tube with a Mo anode, OXFORD XTF5011 (max 50 kV, 1 mA) and a Silicon Drift Detector, thermoelectrically cooled (60EX, Vortex) (FWHM at 160 eV at Fe-Kα), with an active area of 50 mm 2 and a 25 μm thickness Be window. Monochromatization and reduction of the Bremstrahlung from the tube was obtained using a secondary Yttrium target disk (99% purity, 1.0 mm thickness, 25.0 mm diameter, GoodFellow Inc., Huntingdon, UK) in triaxial geometry as it was described in a previous work (Pessanha et al, 2017). The spot size was measured with a radiographic film producing an ellipsoid image with 17 mm × 12 mm.…”
Section: 2portable Setup With Triaxial Geometry -P3xrfmentioning
confidence: 99%
“…All of the energy dispersive X-ray spectra were acquired with the use of a tri-axial setup, which lowers the detection limits of conventional XRF spectrometers through the polarization of the incident beam at a secondary target [ 56 , 57 ].…”
Section: Methodsmentioning
confidence: 99%
“…The X‐ray tube, the secondary target, and the sample are in a triaxial geometry. The background decreases with this arrangement, taking advantage of the effect of the partial polarization of the incident X‐ray beam from the tube and so improving the detection limits . The characteristic radiation emitted by the elements present in the sample was detected by a Si (Li) detector with a 30 mm 2 active area and 8 μm beryllium window.…”
Section: Methodsmentioning
confidence: 99%