A new in situ electrical calibration system for high temperature Calvet calorimeters

Abstract: A new in situ high temperature electrical calibration system was developed at Laboratoire National de Metrologie et d'Essais, Laboratoire Commun de Metrologie and integrated into a heat flux Calvet calorimeter in order to perform accurate and reliable measurements of enthalpy of fusion that are directly traceable to the International System of Units (SI). This system has been designed to enable the calibration of the calorimeter by electrical substitution (Joule effect) as well as the measurement of enthalpy o… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
16
0
1

Year Published

2014
2014
2024
2024

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 14 publications
(17 citation statements)
references
References 13 publications
0
16
0
1
Order By: Relevance
“…La caractérisation métrologique du calorimètre a été réalisée avec ce nouveau système d'étalonnage afin d'étudier l'influence de certains paramètres (énergie, température, durée de dissipation) sur la détermination de la sensibilité des thermopiles. Les résultats de cette caractérisation métrologique ont été publiés par Razouk [12]. Ce moyen de référence et les procédures d'étalonnage et de mesure ont ensuite été utilisés pour la détermination des enthalpies de fusion de l'indium, l'étain et l'argent.…”
Section: Introductionunclassified
“…La caractérisation métrologique du calorimètre a été réalisée avec ce nouveau système d'étalonnage afin d'étudier l'influence de certains paramètres (énergie, température, durée de dissipation) sur la détermination de la sensibilité des thermopiles. Les résultats de cette caractérisation métrologique ont été publiés par Razouk [12]. Ce moyen de référence et les procédures d'étalonnage et de mesure ont ensuite été utilisés pour la détermination des enthalpies de fusion de l'indium, l'étain et l'argent.…”
Section: Introductionunclassified
“…This temperature is defined by the intersection of the interpolated baseline and the tangent drawn at point of inflection of the rising side of the peak. The assumption of the linearity of the sensitivity versus temperature between T1 and T2 was previously verified [12].…”
Section: Calibration and Measurement Proceduresmentioning
confidence: 99%
“…An acquisition system with calibrated devices has been assembled in order to perform accurate and traceable measurements of the energy released in the calorimeter during the calibration process, and of the electromotive force (EMF) delivered by the two thermopiles in both calibration and enthalpy measurement steps. More information on this calibration system has already been published [12]. …”
Section: Figure 1 Instrumented Calibration Cruciblementioning
confidence: 99%
See 1 more Smart Citation
“…Sensitivities from 2.8 µV mW −1 (Setaram HT1000, operation temperature from room temperature to 1000 • C) (Razouk et al, 2013), through 15 µV mW −1 (proprietary instrument with an operating temperature between −73 and 317 • C) (Handy et al, 1993), to 30 µV mW −1 (Setaram C80, operation temperature up to 300 • C) (SETARAM Instrumentation, 2016) can be reached. These data, however, indicate that for TianCalvet calorimeters the maximum operation temperature and the sensitivity behave inversely.…”
Section: The Tian-calvet Calorimeter and Its Special Featuresmentioning
confidence: 99%