A hybrid phantom Monte Carlo-based method for historical reconstruction of organ doses in patients treated with cobalt-60 for Hodgkin's lymphoma

Abstract: Historical radiotherapy treatment plans lack 3D images sets required for estimating mean organ doses to patients. Alternatively, Monte Carlo-based models of radiotherapy devices coupled with whole-body computational phantoms can permit estimates of historical in-field and out-of-field organ doses as needed for studies associating radiation exposure and late tissue toxicities. In recreating historical patient treatments with Co based systems, the major components to be modeled include the source capsule, surrou… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0

Year Published

2019
2019
2023
2023

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(5 citation statements)
references
References 42 publications
0
5
0
Order By: Relevance
“…The compensator‐based IMRT system consists of a cobalt‐60 source and compensator arrays (Figure 1). A commercial cobalt‐60 source with a diameter of 2 cm and a height of 2 cm is designed for this research 15 . The primary collimator is designed to create a 35 × 35 cm 2 field size at 80 cm source to axis distance (SAD).…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…The compensator‐based IMRT system consists of a cobalt‐60 source and compensator arrays (Figure 1). A commercial cobalt‐60 source with a diameter of 2 cm and a height of 2 cm is designed for this research 15 . The primary collimator is designed to create a 35 × 35 cm 2 field size at 80 cm source to axis distance (SAD).…”
Section: Methodsmentioning
confidence: 99%
“…A commercial cobalt-60 source with a diameter of 2 cm and a height of 2 cm is designed for this research. 15 The primary collimator is designed to create a 35 × 35 cm 2 field size at 80 cm source to axis distance (SAD). Two pairs of movable jaws for x and y directions are placed at 31 cm and 37.1 cm from the source, respectively, for adjusting field size.…”
Section: Design Of the Compensator-based Imrt Systemmentioning
confidence: 99%
“…In 2017, Petroccia et al [118] explored the use of the UF/NCI computational phantom library [112] to represent patients historically treated with cobalt-60 gamma-rays for the treatment of Hodgkin’s lymphoma. This patient population is ideal for radiation epidemiology studies of late clinical effects for both secondary cancer and cardiac disease.…”
Section: Applications Of Realistic Computational Human Phantoms In Bimentioning
confidence: 99%
“…In the field of radiotherapy, radiation dose is measured in Gray units (Gy) where 1 Gy is the absorption of 1J of ionizing radiation energy by 1 Kg of tissue [2]; it's described that low doses of radiotherapy (<2 Gy) have been shown to be associated with both secondary cancers and a bit of cardiac toxicity, but true heart damage occurs with radiation dose above 30 Gy [2,3]. In a hybrid phantom Monte Carlo-based method for historical reconstruction of organ doses in patients treated with 60 Co for HL, it was shown that out-fielddoses were 38 % on average.…”
Section: Introductionmentioning
confidence: 99%
“…In a hybrid phantom Monte Carlo-based method for historical reconstruction of organ doses in patients treated with 60 Co for HL, it was shown that out-fielddoses were 38 % on average. Also, distances of 60 cm from the central axis of treatment fields were reported increasing the risk for second malignancies [3].…”
Section: Introductionmentioning
confidence: 99%