1994
DOI: 10.1136/hrt.72.5.486
|View full text |Cite
|
Sign up to set email alerts
|

A comparison of cylindrical and Inoue balloon techniques for mitral valvotomy in patients in the United Kingdom.

Abstract: (Br Heart3' 1994;72:486-491) Several techniques have been described for performing percutaneous mitral balloon valvotomy. Lock et al' used a single cylindrical balloon which reached the mitral valve, via atrial septal puncture, over a guide wire which had been advanced to the aorta. The double cylindrical balloon technique of Al Zaibag et al 2provided a larger dilatation and better haemodynamic results. Catheters with a single shaft and two ("Bifoil") or three ("Trefoil") cylindrical balloons attached have bee… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
5
0

Year Published

1997
1997
2024
2024

Publication Types

Select...
5
1
1

Relationship

0
7

Authors

Journals

citations
Cited by 9 publications
(5 citation statements)
references
References 18 publications
(2 reference statements)
0
5
0
Order By: Relevance
“…In the initial 16 patients, treated before the Inoue balloon was available in Europe, single or double cylindrical balloons were used. We found these two techniques to have the same haemodynamic outcome21 but there was a shorter procedure time and fewer complications with the Inoue technique. The mean (SD) effective balloon dilatation area22 was 3.5 (0.52) cm 2 /m 2 .…”
Section: Methodsmentioning
confidence: 67%
See 1 more Smart Citation
“…In the initial 16 patients, treated before the Inoue balloon was available in Europe, single or double cylindrical balloons were used. We found these two techniques to have the same haemodynamic outcome21 but there was a shorter procedure time and fewer complications with the Inoue technique. The mean (SD) effective balloon dilatation area22 was 3.5 (0.52) cm 2 /m 2 .…”
Section: Methodsmentioning
confidence: 67%
“…15 Echo scoring was by consensus of two observers and was done prospectively in 59 patients and retrospectively from video recordings in the first 21 patients. The latter 53 patients (66%) also had a transoesophageal echo (TOE) study21immediately before the planned procedure. In 17% of these patients left atrial thrombus was detected.…”
Section: Methodsmentioning
confidence: 99%
“…The causes of failure to complete balloon dilatation were failed transseptal puncture (2), intensive vasovagal reaction at transseptal puncture (1), balloon unable to cross the mitral orifice (5), haemopericardium (2), and stroke from dislodged atrial thrombus (1). The success rate for completed valvotomy using the Inoue balloon was 99%, and with cylindrical balloons-used before the Inoue equipment was available in the UK 11 -it was 91%.…”
Section: Methodsmentioning
confidence: 99%
“…14 We found these two methods gave the same haemodynamic result. 11 Right and left heart pressures were recorded using 10 cm above the table as zero. Mean valve gradient was measured by planimetry.…”
Section: Percutaneous Mitral Balloon Valvotomymentioning
confidence: 99%
“…16 Inoue balloon dilatation for rheumatic heart disease was more successful than the over-wire balloon technique. 17 Nevertheless, the suitability of the Inoue balloon for individuals with calcific mitral stenosis is questioned, primarily due to difficulties in attaining optimal left ventricular positioning and concerns regarding mitral valve injury during multiple valve recrossings.…”
Section: Discussionmentioning
confidence: 99%