1979
DOI: 10.3109/00365517909108837
|View full text |Cite
|
Sign up to set email alerts
|

A comparison between the transthoracic electrical impedance method and the direct Fick and the dye dilution methods for cardiac output measurements in man

Abstract: In fourteen patients referred for preoperative evaluation of heart disease we compared the transthoracic electrical impedance method to the direct Fick (four patients) and dye dilution methods (ten patients) for estimation of cardiac output. On statistical analysis of paired differences we found no reliable agreement between the absolute impedance values and the reference methods. The reproducibility was, however, acceptable, and the coefficient of variation for impedance cardiac output was 7.8%, and for imped… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
11
0

Year Published

1981
1981
2016
2016

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 25 publications
(11 citation statements)
references
References 9 publications
0
11
0
Order By: Relevance
“…CO was calculated from the formula: CO=Vi(T,T,)/A, where Vi=volume of indicator, Tb=body temperature and T,=indicator temperature. Corrections for differences in specific heat and density between blood and saline were omitted, in accordance with Enghoff et al (1970). CO (microsphere method) and regional blood flows were calculated by means of a computer program (Schosser et at.…”
Section: Calculations and Statisticsmentioning
confidence: 99%
“…CO was calculated from the formula: CO=Vi(T,T,)/A, where Vi=volume of indicator, Tb=body temperature and T,=indicator temperature. Corrections for differences in specific heat and density between blood and saline were omitted, in accordance with Enghoff et al (1970). CO (microsphere method) and regional blood flows were calculated by means of a computer program (Schosser et at.…”
Section: Calculations and Statisticsmentioning
confidence: 99%
“…cardiac output determined by thermo‐dilution or by Doppler examination ) is time consuming and may not provide sufficient data for meaningful statistical comparisons of the different pacing sites and AV delays in a single patient. This is due to the fact that these techniques can have measurement errors that are in the same order of magnitude than the changes that we are trying to detect . Therefore, they force the use of group statistics to try to get around this technical limitation.…”
Section: Technique For Acute Testingmentioning
confidence: 99%
“…Comparisons of thermodilution cardiac output and metabolic cardiac output have demonstrated statistically significant correlations [7,9,10,11,12,13,14,15], but this does not mean 'agreement' or 'clinical interchangeability'. More recently, a satisfactory agreement has been found between the two methods in stable children [16] and in stable patients with pulmonary hypertension [17].…”
Section: Introductionmentioning
confidence: 99%