“…The CPM setup is a dc-CPM setup in which the light reaching the sample does not pass through a chopper and the photocurrent is detected with an amperemeter. This is opposed to ac-CPM setups where the photocurrent is lock-in detected [6]. In our experiment lock-in detection was used for measuring the intensity of the light.…”
“…The CPM setup is a dc-CPM setup in which the light reaching the sample does not pass through a chopper and the photocurrent is detected with an amperemeter. This is opposed to ac-CPM setups where the photocurrent is lock-in detected [6]. In our experiment lock-in detection was used for measuring the intensity of the light.…”
“…A tungsten-halogen lamp plus grating monochromator were used to scan across the a-Si:H bandgap. The photocurrent was measured in DC mode in order to obtain the optimal image of the localized states distribution [5]. As indications for this distribution and for the defect density, the Urbach slope of the absorption data, respectively the absorption coefficient at 1.2 eV is used.…”
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.