3D silicon microdosimetry and RBE study using 12C ion of different energies

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
21
0

Year Published

2017
2017
2019
2019

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 11 publications
(22 citation statements)
references
References 6 publications
1
21
0
Order By: Relevance
“…The SOI microdosimeter in this study is an array of 3D SVs with size 30 × 30 × 10 μm 3 , fabricated on a high resistivity 10‐μm‐thick n‐SOI active layer, bonded to a 400‐μm‐thick low‐resistivity supporting wafer. The well‐defined 3D SVs has uniform charge collection . SOI microdosimeters, in contrast to the TEPCs, provide high spatial resolution necessary where particle LET varies significantly over a short distance, e.g., toward the end of the BP.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…The SOI microdosimeter in this study is an array of 3D SVs with size 30 × 30 × 10 μm 3 , fabricated on a high resistivity 10‐μm‐thick n‐SOI active layer, bonded to a 400‐μm‐thick low‐resistivity supporting wafer. The well‐defined 3D SVs has uniform charge collection . SOI microdosimeters, in contrast to the TEPCs, provide high spatial resolution necessary where particle LET varies significantly over a short distance, e.g., toward the end of the BP.…”
Section: Methodsmentioning
confidence: 99%
“…Recently, the CMRP has developed and fabricated SOI microdosimeters consisting of micron‐sized well‐defined 3D SVs known as “bridge” microdosimeters. The device was previously characterized in detail and successfully applied in 12 C ion therapy at HIMAC, Japan . Using this microdosimeter, we were able to measure the microdosimetric quantity yD¯ , which was used as an input parameter for the MKM to calculate the RBE 10 corresponding to 10% of human salivary gland (HSG) cell survival.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Calibration of MCA channel number to energy was performed using a calibrated pulse generator. The pulse generator was calibrated to a 300 μm thick planar, silicon, fully depleted PIN diode in response to 5.486 MeV alpha particles from an Am‐241 source . This calibration approach was facilitated by the linear response of the MicroPlus probe.…”
Section: Methodsmentioning
confidence: 99%
“…The derived RBE 10 value at the peak value is 2.6 for the broad‐beam case and it is in a good agreement with the value derived earlier using a Bridge microdosimeter in a water phantom for the same carbon ion beam delivery conditions at HIMAC (NIRS) and is close to the value predicted by the TPS (2.5). Also, this value is slightly higher than the one measured by the TEPC, this is due to the much higher spatial resolution of SOI microdosimeter allowing for the elimination of the volume averaging effect. The peak value of the RBE 10 for spot pencil beam was about 2.4, this is slightly less than the value for the broad‐beam, which could be explained by less contribution of scattering in the distal part of the SOBP.…”
Section: Discussionmentioning
confidence: 72%