Preconcentration procedure using vortex agitator system for determination of trace levels of cadmium by flame atomic absorption spectrometry

Abstract: Um procedimento simples e rápido de microextração líquido-líquido com líquido iônico e agitação vortex (VALLME) foi usado na determinação de quantidades traço de cádmio por espectrometria de absorção atômica com chama (FAAS). O líquido iônico (IL), hexafluorato de 1-hexil-3-metilimidazólio ([Hmim][PF 6 ]), foi usado como solvente extractor. Os íons Cd 2+ foram complexados com pirrolidina ditiocarbamato de amônia (APDC), e então confinados em pequenas gotas de IL pela intervenção do sistema de agitação vortex. … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
3
0

Year Published

2013
2013
2019
2019

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 27 publications
(3 citation statements)
references
References 19 publications
(23 reference statements)
0
3
0
Order By: Relevance
“…The limit of detection (LOD) and quantification were calculated on the basis of 3 S b /m and 10 S b /m (n ¼ 10) (S b is the standard deviation of the blank and m is the slope of calibration curve after extraction), were 0.5 and 1.7 mg/L, respectively. The obtained LOD value for Cd was sufficiently low as compared to a reported work [37] and to be valuable for detecting Cd in different types of water samples.…”
Section: Analytical Figures Of Meritmentioning
confidence: 61%
“…The limit of detection (LOD) and quantification were calculated on the basis of 3 S b /m and 10 S b /m (n ¼ 10) (S b is the standard deviation of the blank and m is the slope of calibration curve after extraction), were 0.5 and 1.7 mg/L, respectively. The obtained LOD value for Cd was sufficiently low as compared to a reported work [37] and to be valuable for detecting Cd in different types of water samples.…”
Section: Analytical Figures Of Meritmentioning
confidence: 61%
“…In this study, FAAS was selected as a determination method and samples were diluted to a certain volume before injecting into the detection system. About vortex agitation, Chamsaz et al [29] firstly employed VALLME for the determination of trace amounts of cadmium by FAAS. In this research, the IL 1-hexyl-3-methylimidazolium hexafluorophosphate ([Hmim][PF 6 ]), was used as an extractant solvent, Cd 2+ was complexed with APDC, and then extracted into fine IL droplets by the assistance of vortex agitator system.…”
Section: Accepted Manuscriptmentioning
confidence: 99%
“…However, this technique also has some disadvantages such as the matrix interfering effects on the analyte signal and higher limit of detection (LOD) than the usual concentrations of metal ions in real samples [6,9]. Therefore, separation and preconcentration steps are often required to achieve accurate, sensitive and reliable results by FAAS [7,10]. Among different preconcentration techniques used in for (ultra) trace analysis of metal ions, solid phase extraction (SPE) is widely used [6,11] because of its flexibility, simplicity, high enrichment factor, high recovery, rapid phase separation, low cost due to low consumption of reagents, and ability of combination with different detection techniques in on-line or off-line modes [12][13][14].…”
Section: Introductionmentioning