Abstract:High-purity germanium detectors are well suited to analysis the radioactivity of samples. In order to reduce the environmental background, low-activity lead and oxygen free copper are installed outside of the probe to shield gammas, outmost is a plastic scintillator to veto the cosmic rays, and an anti-Compton detector can improve the Peak-to-Compton ratio. Using the GEANT4 tools and taking into account a detailed description of the detector, we optimize the sizes of the detectors to reach the design indexes. A group of experimental data from a HPGe spectrometer in using were used to compare with the simulation. As to new HPGe Detector simulation, considering the different thickness of BGO crystals and anti-coincidence efficiency, the simulation results show that the optimal thickness is 5.5cm, and the Peak-to-Compton ratio of 40 K is raised to 1000 when the anti-coincidence efficiency is 0.85. As the background simulation, 15 cm oxygen-free copper plus 10 cm lead can reduce the environmental gamma rays to 0.0024 cps/100 cm 3 Ge (50keV~2.8MeV), which is about 10 -5 of environmental background.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.