Image performance in optoacoustic endoscopy depends markedly on the design of the transducer employed. Ideally, high-resolution performance is required over an expanded depth of focus. Current optoacoustic focused transducers achieve lateral resolutions in the range of tens of microns in the mesoscopic regime, but their depth of focus is limited to hundreds of microns by the nature of their spherical geometry. We designed an ultra-broadband axicon detector with a 2 mm central aperture and investigated whether the imaging characteristics exceeded those of a spherical detector of similar size. We show a previously undocumented ability to achieve a broadband elongated pencil-beam optoacoustic sensitivity with an axicon detection geometry, providing approximately 40 μm-lateral resolution maintained over a depth of focus of 950 μm—3.8 times that of the reference spherical detector. This performance could potentially lead to optoacoustic endoscopes that can visualize optical absorption deeper and with higher resolution than any other optical endoscope today.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.