This paper presents a study for optimal performance of a robot to be used in human surgery (Laparoscope device). He was done in Al-Sader educational hospital in maysan Governorate. The robot was manufactured by Karel Storez company. Connecting linkages were increased to get seven degrees of freedom.The optimal performance was obtained by using genetic algorithm method to choose the optimal path planning in the working area, this was done by making an integrated computer program through MATLAB language (R2013a).The results of best path planning would shorten length without hitting any obstacle, assuming the surrounding environment will be variable, the position and obstacle shapes would be random. We found that the best path planning in every environment depends on objective function. The practical side was made in laboratory of the
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.