Background
Coupling devices have become commonplace in performing venous anastomoses during microvascular free tissue transfer (FTT). However, arterial anastomoses are still most commonly performed using traditional microvascular suture techniques.
Objective
To describe the safety and feasibility of using microvascular coupling devices for free flap arterial anastomosis.
Methods
A systematic review of English language literature was performed for studies that investigated the use of an arterial coupler for microvascular FTT in human patients. A comprehensive search of MEDLINE (January 1948 to August 2018), EMBASE (January 1974 to August 2018), and Web of Science was performed.
Results
Fifteen studies were included. All studies were retrospective case series. A combined total of 395 arterial anastomoses were attempted with a coupling device. All studies except for one used the 3M Unilink/Synovis coupling device. One study used a novel absorbable coupling device. The coupling device was aborted and converted to a suture technique in 8.4% of attempted anastomoses. Rupture of the anastomotic device was reported in only 1 patient (0.3%). Thrombosis was also infrequent at 1.9%. The quality assessment showed a high risk of bias in all studies.
Conclusion
In selected patients, coupling devices for arterial anastomoses have a good success rate with low rates of thrombosis based on limited quality evidence.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.