Background:We developed a method, called the metabolic heat conformation (MHC) method, for the noninvasive measurement of blood glucose. The MHC method involves the measurement of physiologic indices related to metabolic heat generation and local oxygen supply, which correspond to the glucose concentration in the local blood supply. Results: Regression analysis was performed to compare the noninvasive method with the hexokinase method, using 127 data points (109 data points from diabetic patients, 18 data points from nondiabetic patients) with glucose concentrations ranging from 3.0 to 22.5 mmol/L (54 -405 mg/dL). The correlation coefficient (r) was 0.91. Reproducibility was measured for healthy fasting persons; the CV was 6% at 5.56 mmol/L (100 mg/dL). Conclusions: These data provide preliminary evidence that the MHC method can be used to estimate blood glucose concentrations noninvasively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.