Background
The authors developed a cross‐laser projection system (CLP) to place a femoral neck‐sparing short stem using the minimally invasive anterolateral supine approach in total hip arthroplasty. This study aimed to verify the utility of CLP.
Methods
Thirty joints were assessed with the MiniHip (Corin). The authors compared femoral component implantation with a patient‐specific femoral osteotomy guide (PSG) for the femoral neck‐cut (PSG group), with the CLP attached to the rasp handle to irradiate the cross‐laser to the target of PSG (CLP group), and without PSG or CLP (control group).
Results
In the CLP group, the positional deviation of anteversion, anterior/posterior tilt and varus/valgus placement of the stem postoperatively were 1.8° ± 0.2°, 2.0° ± 2.0° and 2.0° ± 0.1°, respectively. The positional deviation of anteversion (p < 0.001) and anterior/posterior tilt (p = 0.036) were significantly smaller than those in the other groups.
Conclusions
CLP improves the accuracy of MiniHip femoral prosthesis placement.
Aims Although the short stem concept in hip arthroplasty procedure shows acceptable clinical performance, we sometimes get unexplainable radiological findings. The aim of this retrospective study was to evaluate changes of radiological findings up to three years postoperatively, and to assess any potential contributing factors on such radiological change in a Japanese population. Methods This is a retrospective radiological study conducted in Japan. Radiological assessment was done in accordance with predetermined radiological review protocol. A total of 241 hips were included in the study and 118 hips (49.0%) revealed radiological change from immediately after surgery to one year postoperatively; these 118 hips were eligible for further analyses. Each investigator screened whether either radiolucent lines (RLLs), cortical hypertrophy (CH), or atrophy (AT) appeared or not on the one-year radiograph. Further, three-year radiographs of eligible cases were reviewed to determine changes such as, disappeared (D), improved (I), stable (S), and progression (P). Additionally, bone condensation (BC) was assessed on the three-year radiograph. Results CH was observed in 49 hips (21.1%), AT was observed in 63 hips (27.2%), and RLLs were observed in 34 hips (14.7%) at one year postoperatively. Among 34 hips with RLLs, 70.6% showed change of either D or I on the three-year radiograph. BC was observed in younger patients more frequently. Conclusion The Fitmore stem works well in a Japanese population with favourable radiological change on hips with RLLs. Longer-term follow-up is required to determine clinical relevance. Cite this article: Bone Jt Open 2022;3(1):20–28.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.