We describe a stacked detector made of thin cadmium telluride (CdTe) diode detectors. By using a thin CdTe device, we can overcome the charge loss problem due to the small mobility and short lifetime of holes in CdTe or cadmium zinc telluride (CdZnTe) detectors. However, a CdTe detector with a thickness of more than 5 mm is needed for adequate detection efficiency for gamma-rays of several hundred keV. Good energy resolution and good peak detection efficiency are difficult to obtain using such a thick CdTe detector. The stacked detector enabled us to realize a detector with both high-energy resolution and good efficiencies for gamma rays up to several hundred keV. In order to verify this concept, we constructed a prototype made of ten layers of a 0.5-mm-thick CdTe diode detectors with a surface area of 21.5 mm 21.5 mm. With this, we have achieved 5.3-keV and 7.9-keV energy resolution [full width at half maximum (FWHM)] at 356 keV and 662 keV, respectively, at the temperature of 20 C.Index Terms-Cadmium telluride (CdTe), CdTe diode detector, cadmium zinc telluride (CdZnTe), stacked detector.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.