This paper investigates a quality assurance (QA) phantom specially designed to verify the accuracy of dose distributions and monitor units (MU) calculated by clinical treatment planning optimization systems and by the Monte Carlo method for intensity-modulated radiotherapy (IMRT). The QA phantom is a PMMA cylinder of 30 cm diameter and 40 cm length with various bone and lung inserts. A procedure (and formalism) has been developed to measure the absolute dose to water in the PMMA phantom. Another cylindrical phantom of the same dimensions, but made of water, was used to confirm the results obtained with the PMMA phantom. The PMMA phantom was irradiated by 4, 6 and 15 MV photon beams and the dose was measured using an ionization chamber and compared to the results calculated by a commercial inverse planning system (CORVUS, NOMOS, Sewickley, PA) and by the Monte Carlo method. The results show that the dose distributions calculated by both CORVUS and Monte Carlo agreed to within 2% of dose maximum with measured results in the uniform PMMA phantom for both open and intensity-modulated fields. Similar agreement was obtained between Monte Carlo calculations and measured results with the bone and lung heterogeneity inside the PMMA phantom while the CORVUS results were 4% different. The QA phantom has been integrated as a routine QA procedure for the patient's IMRT dose verification at Stanford since 1999.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.