Background The aim of this study was to develop and validate accurate and precise UPLC method with tandem mass spectrometry (Waters) for the determination of bexarotene in human plasma using bexarotene D4 as internal standard (IS). Results The retention time of bexarotene was 2.75 ± 0.30 min. The method was validated with respect to system suitability, linearity, accuracy, precision, matrix effect, auto sampler carryover test, and recovery. Linearity was found to be 1.04 to 351.93 μg/mL. LOQQC, LQC, INTQC, MQC, and HQC were found to be 1.0550, 2.7800, 25.2700, 131.61, and 263.23 respectively. The mean percentage recovery was found to be 95.72% Conclusion The bioanalytical method, a selective and sensitive liquid chromatography-mass spectrometry method to quantitate bexarotene in K2EDTA human plasma over the concentration range 1.0440 to 351.9320 ng/mL, was successfully validated. This method is suitable for sample analysis to support bioequivalence/bioavailability and/or pharmacokinetic studies involving formulations of bexarotene.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.