Background Recent indirect evidence of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) transmission during endoscopic endonasal procedures has highlighted the dearth of knowledge surrounding aerosol generation with these procedures. As we adapt to function in the era of Coronavirus Disease 2019 (COVID-19) a better understanding of how surgical techniques generate potentially infectious aerosolized particles will enhance the safety of operating room (OR) staff and learners. Objective To provide greater understanding of possible SARS-CoV-2 exposure risk during endonasal surgeries by quantifying increases in airborne particle concentrations during endoscopic sinonasal surgery. Methods Aerosol concentrations were measured during live-patient endoscopic endonasal surgeries in ORs with an optical particle sizer. Measurements were taken throughout the procedure at six time points: 1) before patient entered the OR, 2) before pre-incision timeout during OR setup, 3) during cold instrumentation with suction, 4) during microdebrider use, 5) during drill use and, 6) at the end of the case prior to extubation. Measurements were taken at three different OR position: surgeon, circulating nurse, and anesthesia provider. Results Significant increases in airborne particle concentration were measured at the surgeon position with both the microdebrider (p = 0.001) and drill (p = 0.001), but not for cold instrumentation with suction (p = 0.340). Particle concentration did not significantly increase at the anesthesia position or the circulator position with any form of instrumentation. Overall, the surgeon position had a mean increase in particle concentration of 2445 particles/ft3 (95% CI 881 to 3955; p = 0.001) during drill use and 1825 particles/ft3 (95% CI 641 to 3009; p = 0.001) during microdebrider use. Conclusion Drilling and microdebrider use during endonasal surgery in a standard operating room is associated with a significant increase in airborne particle concentrations. Fortunately, this increase in aerosol concentration is localized to the area of the operating surgeon, with no detectable increase in aerosol particles at other OR positions.
Objective
Recent anecdotal reports and cadaveric simulations have described aerosol generation during endonasal instrumentation, highlighting a possible risk for transmission of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS‐CoV‐2) during endoscopic endonasal instrumentation. This study aims to provide a greater understanding of particle generation and exposure risk during endoscopic endonasal instrumentation.
Study Design
Prospective quantification of aerosol generation during office‐based nasal endoscopy procedures.
Methods
Using an optical particle sizer, airborne particles concentrations 0.3 to 10 microns in diameter, were measured during 30 nasal endoscopies in the clinic setting. Measurements were taken at time points throughout diagnostic and debridement endoscopies and compared to preprocedure and empty room particle concentrations.
Results
No significant change in airborne particle concentrations was measured during diagnostic nasal endoscopies in patients without the need for debridement. However, significant increases in mean particle concentration compared to preprocedure levels were measured during cold instrumentation at 2,462 particles/foot
3
(95% CI 837 to 4,088;
P
= .005) and during suction use at 2,973 particle/foot
3
(95% CI 1,419 to 4,529;
P
= .001). In total, 99.2% of all measured particles were ≤1 μm in diameter.
Conclusion
When measured with an optical particle sizer, diagnostic nasal endoscopy with a rigid endoscope is not associated with increased particle aerosolization in patient for whom sinonasal debridement is not needed. In patients needing sinonasal debridement, endonasal cold and suction instrumentation were associated with increased particle aerosolization, with a trend observed during endoscope use prior to tissue manipulation. Endonasal debridement may potentially pose a higher risk for aerosolization and SARS‐CoV‐2 transmission. Appropriate personal protective equipment use and patient screening are recommended for all office‐based endonasal procedures.
Level of Evidence
3
Laryngoscope
, 2020
Recognizing the value of including complex pyridines in small-molecule screening collections, we developed a previously unexplored [2 + 2 + 2]-cycloaddition of silyl-tethered diynes with nitriles. The tether provides high regioselectivity, while the solvent THF allows catalytic CpCo(CO)2 to be used without exogenous irradiation. One of the resulting bicyclic and monocyclic (desilylated) pyridines was identified as an inhibitor of neuregulin-induced neurite outgrowth (EC50 = 0.30 µM) in a screen that probes a pathway likely to be involved in breast cancers and schizophrenia.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.