Overall, only a small number of tube sizes (14Fr and 20Fr) with selected formulas revealed a significant difference between ENFit and Legacy tubes, with remaining studies finding no significant difference. Regression analysis revealed that variables such as formula, size of tube, blender used, and time of blending may have more impact on compression force.
Overall, for the low-profile, 18F, and 20F sizes, the ENFit tubes had similar flow rates when compared with the legacy tubes. For the 14F and 24F sizes, the flow rate of ENFit tubes was significantly lower, which could result in longer EN delivery for patients who are using these tubes to provide gravity feeding.
There is wide variability in the flow dynamics in ENFit connectors with significant potential impact on many facets of HEN, including medicine delivery, blenderized feeds, venting, and compliance with EN due to increased time to administer feeds. We highly recommend additional testing of flow dynamics, including gravity flow, as ENFit tubes are being developed and adopted.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.