ObjectiveThe aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010.Participants and methodsThe evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97–190), diastolic BP was 90.9±12.9 mmHg (range: 68–120), and arm circumference was 30.8±4 cm (range: 25–38.5).ResultsThe device successfully fulfilled the established criteria of the validation protocol. The device overestimated systolic BP by 0.6±5.7 mmHg and diastolic BP by 2.2±5.1 mmHg. The specially designed cuff HEM-FL1 to cover a broad range of arm circumferences and self-placement fulfilled the requirements of the International Protocol.
Reliability of an automatic monitor for blood pressure measurement for DBP. The RC for SBP (6.2, 5.2 and 5.8 mmHg) and DBP (4.7, 4.2 y 5.2 mmHg) was similar between the observers and the device. The ICC for SBP was 0.990 (95% CI 0.983 to 0.995, p < 0.01) and 0.986 (95% CI 0.977 to 0.991, p < 0.01) (Rev Med Chile 2018; 146: 190-195)
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.